A&D Monitor Troubleshooting

If you haven't set up your A&D monitor, please follow the instructions on the A&D Setup page.

If you see an error, remove batteries for 5–10 minutes, then re-insert batteries. If that doesn’t work, follow the tips below or check your monitor’s instruction guide.

Symbol What it means What to do
Appears while measurement is in progress. It blinks when the pulse is detected. Measurement is in progress. Remain as still as possible.
Battery Charge The battery power indicator during measurement. When battery is low, replace with new battery.
Err Unstable blood pressure due to movement during measurement. Take another measurement and remain still.
Err The systolic and diastolic values are within 10 mmHg of each other. Apply the cuff correctly and take another measurement.
Err The pressure value did not increase during the inflation. Apply the cuff correctly and take another measurement.
Err CUF The cuff was not applied correctly. Apply the cuff correctly and take another measurement.
E PUL. DISPLAY ERROR The pulse is not detected correctly. Apply the cuff correctly and take another measurement.
Err E or Err 9 Device internal error. Remove the batteries, press the power button, then install the batteries again. If you still see the error, contact us.
Err 10 or Err 11 Pairing has not been performed correctly. Remove and re-insert the batteries. Try pairing again.

If your electronic transmissions are not working, you can follow these instructions for manual reporting.